Toward automated intraocular laser surgery using a handheld micromanipulator - Robotics Institute Carnegie Mellon University

Toward automated intraocular laser surgery using a handheld micromanipulator

Conference Paper, Proceedings of (IROS) IEEE/RSJ International Conference on Intelligent Robots and Systems, pp. 1302 - 1307, September, 2014

Abstract

This paper presents a technique for automated intraocular laser surgery using a handheld micromanipulator known as Micron. The novel handheld manipulator enables the automated scanning of a laser probe within a cylinder of 4 mm long and 4 mm in diameter. For the automation, the surface of the retina is reconstructed using a stereomicroscope, and then preplanned targets are placed on the surface. The laser probe is precisely located on the target via visual servoing of the aiming beam, while maintaining a specific distance above the surface. In addition, the system is capable of tracking the surface of the eye in order to compensate for any eye movement introduced during the operation. We compared the performance of the automated scanning using various control thresholds, in order to find the most effective threshold in terms of accuracy and speed. Given the selected threshold, we conducted the handheld operation above a fixed target surface. The average error and execution time are reduced by 63.6% and 28.5%, respectively, compared to the unaided trials. Finally, the automated laser photocoagulation was demonstrated also in an eye phantom, including compensation for the eye movement.

BibTeX

@conference{Yang-2014-120589,
author = {S. Yang and R. A. MacLachlan and C. N. Riviere},
title = {Toward automated intraocular laser surgery using a handheld micromanipulator},
booktitle = {Proceedings of (IROS) IEEE/RSJ International Conference on Intelligent Robots and Systems},
year = {2014},
month = {September},
pages = {1302 - 1307},
}