Characteristics of hand motion of eye surgeons
Conference Paper, Proceedings of 19th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC '97), pp. 1690 - 1693, October, 1997
Abstract
Overall manual accuracy and motion frequency in simulated microsurgery have been studied. Eye surgeons were tested in two tasks: attempting to hold an instrument still, and repeatedly actuating. The RMS error and overall motion range were measured. Spectral analysis was also performed. The average RMS errors for the two test conditions were 49 /spl mu/m and 133 /spl mu/m, respectively, and the average range of motion while trying to hold still was 202 /spl mu/m. Substantial low-frequency motion occurred under both conditions. On average, 98.9% of the total power of voluntary movements tested was found to be below 2 Hz.
BibTeX
@conference{Riviere-1997-14488,author = {Cameron Riviere and R. Scott Rader and Pradeep Khosla},
title = {Characteristics of hand motion of eye surgeons},
booktitle = {Proceedings of 19th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC '97)},
year = {1997},
month = {October},
pages = {1690 - 1693},
}
Copyright notice: This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author's copyright. These works may not be reposted without the explicit permission of the copyright holder.